BRIGHT STAR TEST TARGETS
We would like to investigate
what the possibilities are for broadening the scope of WET
targets to include other stars with potential for asteroseismology.
Amongst these are the rapidly oscillating Ap (roAp) stars. Because these
stars are much brighter than white dwarfs, there are some technical
challenges which must be addressed before committing to observing a roAp
star. So the following procedure describes a test to assess the stability
of your photometer and the sky above your observatory to see whether roAp
stars can be observed with your setup.
Why should there be doubts? The answer is that in observations of
white dwarfs, a multitude of small instabilities may be hidden beneath
the photon noise from these faint stars. Rapidly oscillating Ap
stars are much brighter so photon noise is almost negligible.
Also, they are *much* lower amplitude (1 per cent peak-to-peak is a
large amplitude star), so instrumental and atmospheric stability are of
utmost importance.
Please find below links to 4 stars. Each link contains co-ordinates and a
finding chart:
Bright stars for smaller telescopes with filters:
Fainters stars for larger telescopes (especially if unfiltered):
Please observe one of these targets. A B filter is preferred, but if
you don't have one, a V filter will be ok. Smaller telescopes (1-m) with
a B filter should shoot for one of the first 2 stars, larger
telescopes (2-m) for one of the last 2 stars. One of each pair is
accessible in the evening, the other in the morning.
Try and choose a star that gives you about a factor 2.5 smaller than the count
rate limit of your tube (e.g. if your tube is limited to a million counts
per sec, try and choose a target that will give you as near as you can
get to 400 000 counts per sec). Obviously, where this is not possible,
choose a fainter or brighter target. Try not to go below 100 000
counts per sec if you can avoid it. If you do, photon and sky noise
will be detectable and the test will not be as sensitive. It is, in
fact, best if you can stay above 250 000 counts per sec.
(If you don't know the count rate limit
of your tube, be sure to find it out before you try this as you could
damage your tube. A million counts/sec is a typical limit but some
red-sensitive GaAs tubes have lower limits, so be careful).
Observe only if you are sure it is photometric, and guide as carefully
as you can. You need to get millimag accuracy data, from one point
to the next.
Get one hour of data with 10-sec integrations. Use a 30 arcsec aperture,
not less. Use a larger aperture if you consider it necessary. Sky
background is not a problem here so your normal instincts to choose
the smallest aperture consistent with losing no light do not apply. The
detectable skirt on the point spread function of stars this bright
extends *much* further than for white dwarfs, and you need to avoid
"clipping off" even 0.0001 of the light. Therefore using a larger
aperture than you are accustomed to lets in more sky, but that is
less important than making sure you lose no detectable starlight.
Send the data you get to HQ as usual. Please send information about the
setup *along* with the data file so this information and the data file are
always together. The setup information needed includes: (1) the telescope;
(2) the kind of photometer; (3) the filter used; (4) the photomultiplier
tube you used; (5) the target star; (6) the observer.
Best of luck...
PS. We would also like you to measure the dead time of your
photomultiplier tubes because with stars as bright as this,
dead time count losses are significant. Remarkably, variations
of dead time losses as the star's brightness changes between
maximum and minimum are detectable. Thus, an accurately
known dead time is important. Here are two
procedures to determine deadtime:
Darragh O'Donoghue dod@saao.ac.za.
Last updated: 28-Oct-1999.