
Xcov 20 Deadtime Measurement
Deadtime Measurement Number 1
- . Using the Texas photometer as a basis, choose two apertures
(one smaller, one larger) and arrange to illuminate them both
with a light source that shows stability of counting rate for
several minutes. I found an LED that had been on for long
enough to come to temperature equilibrium works fine, if it
is powered by a regulated power supply.
- . Take two readings of the counting rate, one through each
aperture, the higher chosen so
that the dead time should be significant at that counting
rate -- between 500,000 cps and 2,000,000 cps worked for me.
- . Now decrease the illumination intensity, so the counting rate
drops about a factor of 10 or more. After the LED is stable
again, take two more readings, through each of the same
apertures.
- . There are four data points (the readings) and four unknowns:
- The ratio of the aperture areas
- The ratio of the lower to the higher intensity
- The "true" counting rate if the detector were ideal
- The dead time
- I worked out the equations for this, but I don't have them
in front of me ... I can probably find them somewhere. But
the exercise is not very hard, and is useful to know about.
I recommend, as an exercise to the student, that you work out
the simultaneous equations and see if you get something you
can understand ... and believe in. They might even be the same
as the ones I worked out many years ago ...
To first order, if the dead time is negligible at the lower
counting rate, then the observed ratio calibrates the aperture
area ratio, which should be the same at the higher rate. Any
discrepancy can be blamed on the dead time. The proper way
to do it, though, is to solve the simultaneous equations for
the dead time explicitly. I got the 20 ns dead time figure
for the current system that way. Repeated measurements gave
the same dead time to within 10 percent.
ADVANTAGES: Takes little time - Can be done any time
DISADVANTAGES: Less accurate - Sensitive to measurement errors
- Requires constant artificial light source
Deadtime Measurement Number 2
- On a nice clear evening, wait until the Sun is about 8 degrees
below horizon.
- Find an aperture which gives you a count rate of about
1,000,000 cps at a blank piece of sky. Use a filter, if useful.
ALWAYS start with the smallest aperture when trying to approach
this count rate.
- Start recording data. Switch back and forth between this aperture
and a second one with at most a quarter of the other's diameter;
use 2x10 seconds of integration in each aperture. Continue
switching until sky counts drop below about 3000 cps in the larger
aperture. Then do longer integrations in both apertures to get good
statistics for measuring the aperture size ratio.
- Take a number of dark integrations, then quit this run.
- Check that there is no star in the larger aperture.
- To determine the deadtime, first subtract the mean dark count
from the data. Separate the curves for the different apertures.
- Determine the aperture area ratio from the count rate ratio in
the last (flat) part of the curves when you used longer integration
times to get better photon statistics. Multiply the curve with the
smaller count rates by this ratio.
- Bring the steep parts of the curves to agreement by using trial
deadtimes, choose the best value. Small adjustments to the aperture
area ratio may be necessary.
Or plot a graph of ratio of counts in big to (counts in small
interpolated
to time of counts in big) versus counts in big. This will be a straight
line whose negative slope will be (1-alpha)tau where alpha is the intercept
= ratio of apertures and tau is the dead time in seconds (convert counts
per 10 sec to counts per sec before plotting).
- Obviously, you can also do this in morning twilight in reverse order.
ADVANTAGES: More accurate
- Incorrect measurements are easily recognized
DISADVANTAGES: Might take away precious observing time
- Requires clear skies
Send comments to wetmaster.
Last updated on 6 Nov 2000.